|
FlowJo gives you the ability to compensate your
data. This may be necessary in cases where the compensation was
inappropriately set during sample collection (although if the sample
was over-compensated, then there is no recourse). Also, there may
be cases where the instrument is not capable of compensating between
certain channels (for instance, to correct for the spillover between
fluorescein and Cy5PE).
For a description of why compensation is necessary,
the underlying concepts behind compensation, and some of the pitfalls
of improper compensation, see the "Compensation:
A perspective" by Mario Roederer (you
need to be connected to a network to view this site). The remainder
of the documentation below is NOT meant to teach you about compensation,
simply how to use FlowJo to perform that function.
FlowJo computes the compensation matrix on control samples much
the way you would manually set the compensation. To do this, you
will select gates on positive and negative populations for each
of these stain, and tell FlowJo to calculate the compensation matrix
based on these stains.
Thus, you must collect samples that are singly-stained (as
well as unstained) for each of the fluorescences that require compensation.
Ideally, you would use a stain that only labels a portion of the
sample population, so that you have an unstained set of cells in
each tube. It is important to remember that the negative population
and positive population must share the same autofluorescence characteristics
for compensation to be valid (i.e., don't use monocytes as a negative
control for a lymphocyte stain; if you are compensating a stain
on fibroblasts, use an unstained fibroblast control).
Steps involved in compensation by FlowJo:
- Define positive and negative gates for each fluorescence channel
requiring compensation
- Open the compensation
window; drag the positive and negative
populations into the appropriate boxes
- Compute and save the matrix
- Apply the matrix to the appropriate samples
If you need to generate another compensation matrix for other samples
in the experiment, you can just repeat steps 2 through 4 as needed.
You may also view an example
of the entire compensation process, as
a tutorial.
Once you have defined a compensation matrix, it is saved as a file
on your Macintosh as well as in the workspace itself. You may subsequently
apply that compensation matrix to other samples in the same workspace
by selecting it from the menu; you can use that matrix in other
workspaces by reading it in from the file that you saved. Note that
a compensation matrix is generally valid only for samples collected
during a single collection run!
Any sample that has been compensated is marked with a blue bar
next to the sample name in the workspace window. Compensated samples
have new parameters added to their list: for each fluorescence channel
to be compensated, a new parameter is created. The parameter name
is bracketed with "<>": e.g., when FITC and PE are compensated
against each other, two new parameters named "<FITC>" and
"<PE>" will be created. Remember to select these new parameters
in the graph or statistics windows!
Finally, note that you can increase the dynamic
range of compensated parameters (at the low end of intensity) by
a setting in the Preferences.
This can improve the visualization of compensated parameters, especially
if a lot of compensation is being performed.
You may also click on the topics below to get more help on:
|