|
FlowJo has a unique platform that allows
you to calibrate any collected parameter. This gives you the ability
to express any parameter's values (graphical display, any statistical
calculations) in the units of your choice. Most commonly, this is
used to convert the scaling into absolute number of molecules (given
a standard that converts between the fluorescence intensity collected
on your instrument and absolute numbers of fluorophores).
Once you have created a calibrated parameter, it
is added to your sample as if it were just another parameter. You
can apply the calibration standard to any other sample in your collection
(and you can apply the calibration to an entire group, if you wish).
You may have as many different calibration standards in a workspace
as you want; in fact, you can even apply multiple calibration standards
to the same sample (a new parameter will be added for each calibration
that you apply).
When you graph your data, select the calibrated
parameter on the axis of choice--the scale values on the axis are
now drawn in your calibrated units! For example, if you calibrate
to absolute number of molecules per cell, then you can display these
numbers on the axis. In addition, any statistics you calculate (medians,
CVs, etc.) will be expressed in units of absolute molecules. Finally,
you can even enter a gate based on those units: you could, for example,
specify a gate on all cells with more than 5,000 molecules of an
antigen.
There are two methods by which you can calibrate
a parameter: manual or reference. Use the manual mode when you want
to enter the exact conversion values yourself. Use the reference
when you have collected a standard which you can use to convert
to your units (for example, a commercial bead set where each bead
has a defined absolute intensity; or, a sample stained with an antibody
that has a known binding capacity to the cells). When you use a
reference, FlowJo automatically finds all of the peaks in the fluorescence,
and asks you to enter any or all of the calibration values for those
peaks. FlowJo then performs a linear (or logarithmic) least squares
regression to determine the best calibration value, and creates
the calibrated parameter.
A "Calibration standard" is FlowJo's internal
representation of the conversion of fluorescence to absolute units.
Calibration standards are stored in the workspace, permanently.
They are accessible as Derived parameters; thus, if you want to
apply a Calibration Standard to another sample (after defining it),
select the appropriate derived parameter under the "Derived
Parameters" menu item. For more information on derived parameters,
click here.
To define a calibration standard, you will be presented
with the "Define New Calibration
Standard" dialog window" that asks you to define which
parameter you want to calibrate, and then which method to use. This
dialog window is shown when you select "Calibrate..."
under the Workspace menu. To see how to use either method to create
a new Calibration Standard, click on the link to this dialog window.
In addition, you can see two examples of creating
Calibration Standards. Both use reference
samples; one uses a bead set, the other, the expression of CD4 on
lymphocytes as a defined set point. Examine these two series of
pages to see how easy it is to calibrate your samples with FlowJo!
Download a Calibration Workspace
with Demo Data to try out this platform.
|