 Compensation Interface
The graphic below shows what the compensation definition window looks like and the controls. To more fully understand this window, try the compensation tutorial.

 

There are several ways to apply the positive and negative populations you have defined into the comp matrix dialog boxes. The simplest is to define a positive and negative population on each single stained comp control sample and drag each of these to the boxes beside the appropriate parameter -see (a) below. However, you can have a "universal" negative that can be used for all compensation controls that only have positive populations. For example, when using antibody capture beads to compensate--each sample has beads that captured one color (with no negatives). Another sample contains unstained beads and is used as the negative population for each parameter - see (d) below.

a) Drag the positive and negative subsets from the workspace to the boxes beside each parameter in the Comp Definitions window.

OR

b) Drag a subset that has one or two gated subsets to the Auto Assign Button. FlowJo looks at each subset you drop on it, and looks for child gates. If there is only one child gate, it assumes that this represents a positive comp control populations. If there are two child gates, it assumes that there is a positive and a negative.

OR

c) Drag a sample that contains a mix of singly stained comp control bead samples to the PolyComp Button. FlowJo automatically finds the positive populations for each parameter. Define the negative population by dropping the gated unstained bead sample onto the universal negative button.

OR

d) Drag an unstained sample or negative subset to the Universal Negative button. FlowJo puts this negative population into all the negative boxes for each parameter. In general, you should drop the universal negative on the comp window first--because if you do it second, it would replace any existing negative populations you already added.

If you have dragged the incorrect population into one of the boxes, simply select it and hit the delete key to remove it from the box. Or, you can drag and drop the correct population onto the box, replacing the existing one. You can know what population is in a box by selecting that box (clicking on it); the name of the population that you dragged into the box will be shown in the lower left corner of the window.

Note that some flow cytometers collect a fluorescence parameter in the SSC-W channel. This channel will therefore be displayed in the compensation dialog window when sample files in the workspace contain this parameter.

Go back to the overview of compensation.