| FlowJo
for Windows
FlowJo gives you the ability to compensate your
data. This may be necessary in cases where the compensation was
inappropriately set during sample collection (although if the sample
was over-compensated, then there is no recourse). Also, there may
be cases where the instrument is not capable of compensating between
certain channels (for instance, to correct for the spillover between
FITC and Cy5PE).
For a description of why compensation is necessary,
the underlying concepts behind compensation, and some of the pitfalls
of improper compensation, see the "Compensation:
A perspective" by Mario Roederer
(internet connection required.) The remainder of the documentation
below is NOT meant to teach you about compensation, simply how to
use FlowJo to perform that function.
FlowJo computes the compensation matrix on control
samples much the way you would manually set the compensation. To
do this, you will create gates on positive and negative populations
for each of these stain, and tell FlowJo to calculate the compensation
matrix based on these stains.
The matrix files are saved to a user specified directory
whenever a matrix is defined. The files have an .MTX extension and
can be edited with an XML or a text editor (for example notepad,
wordpad..). Please note editing the matrix file is not recommended!
To perform software compensation, you must collect
samples that are singly-stained (as well as unstained) for
each of the fluorescences that require compensation. Ideally, you
would use a stain that only labels a portion of the sample population,
so that you have an unstained set of cells in each tube. It is important
to remember that the negative population and positive population
must share the same autofluorescence characteristics for compensation
to be valid (i.e., don't use monocytes as a negative control for
a lymphocyte stain; if you are compensating a stain on fibroblasts,
use an unstained fibroblast control).
Steps involved in compensation by FlowJo are outlined
in the Compensation Tutorial.
Any sample that has been compensated is marked with
a bar next to the sample name in the workspace window. Compensated
samples have new parameters added to their list: for each fluorescence
channel to be compensated, a new parameter is created. The parameter
name is labeled with "comp-": e.g., when FITC and PE are
compensated against each other, two new parameters named "Comp-FITC"
and "Comp-PE" will be created. Remember to select these
new parameters in the graph or statistics windows!

Download a Compensation Demo
Data to try out this platform.
|