| If you choose the manual calibration
method for calibrating a parameter, you will be shown the following
window (with the exception that the "Change Relative Scaling"
checkbox is not selected, and the second calibration point entry box
is grayed out).
For a simple calibration, you can just enter one
pair of values in the "Reference" box. Enter a known absolute
fluorescence in the leftmost box, and then enter the corresponding
channel number OR the corresponding scale value (of the original
parameter) in the boxes on the right. If this is all you enter,
then FlowJo simply rescales the original parameter by an appropriate
multiplier so that the reference position is correct.
The scale value (or channel value) that you enter
might be derived from a gating or calculation on the parameter from
this or another sample that you have performed manually. For example,
you may have stained cells with anti-CD4, and determined that the
mean fluorescence of the CD4 positive cells is 25.6 scale units.
You would then enter 25.6 in the "Scale Value" box, and
50,000 in the known value box (there are 50,000 molecules of CD4
per CD4 T cell), and then click on "Define Parameter".
FlowJo creates a new parameter which is absolute molecules based
on the conversion you entered.
Note that the one-point standard curve is the same
as fitting a straight line through the origin (on a linear scale
parameter), or maintaining the same dynamic range as the original
parameter (for a log scale parameter). In the latter case, this
means that the number of decades for the scaled parameter is identical
to the number of decades for the original collected parameter.
If you wish to perform a 2-point standard curve,
you can check the box "Change Relative Scaling". For linear
parameters, you will then enter the scale value (or channel number)
corresponding to Zero absolute units. For log parameters, in which
there can be no zero, you will enter the scale value (or channel
number) corresponding to 1 absolute unit. The net effect of this
is a complete rescaling of the parameter. For linear parameters,
this may "move" the events left or right on the graph,
because zero absolute units will always be the left-most value on
the graph.
For log parameters, rescaling means that you may
change the number of decades in the parameter (in order to fit both
known points). This means that the absolute values may no long scale
with fluorescence intensity, assuming that the original parameter
was a reasonable estimate of that intensity.
The effect of 2-point calibrations is too complex
for the scope of this help documentation. If you need more help
understanding 2 point calibrations, please contact
the FlowJo administrators for more information. |